PowerMax-USB LM-150 LS

PowerMax-USB LM-150 LS

SKU 1275678
UOM Each

USB-enabled, compact laser power sensor for intermittent use from 100 mW to 150W. Beam position sensing, 0.25 to 10.6 µm spectral range, 19 mm diameter active area.


Price $2,480.00 USD


Details

Compact convectively cooled thermopile, with direct USB interface, for measuring pulsed & CW lasers intermittently up to 150W with beam position measurement:

2 minutes at 150W

5 minutes at 100W

20 minutes at 40W

 

Key Features:

·      100 mW to 150W range (intermittent)

·      19 mm diameter active area

·      Beam position measurement

·      Spectrally flat broadband coating

·      0.25 to 10.6 µm spectral range

·      Conductively cooled

·      Fiber adapters available

·      Direct USB interface

·      ISO 17025 accredited

This PowerMax-USB LM-150 LS sensor is terminated in a USB cable that interfaces directly to a PC. A DB25 model (PN 1098452) is available.

This LM-150 LS version has a rectangular form factor, whereas the LM-150 FS (PN 1098394) is round.

SMA fiber adapter (PN 1098589) and FC/PC fiber adapter (PN 1098339) accessories are available when used with 1.035-40M Adapter Ring (PN 33-9432-000).

These unique position sensing thermopiles display the position of the incident beam on the sensor disk, in addition to measuring power. This information allows the user to safely and precisely align the sensor and the beam being measured. This is particularly useful for non-visible beams.

Resources

Datasheets
Owner's Manuals
Owner's Manuals
Drawings
Software
Software
Software

Specs

General

Cable Type
USB
Cooling Type
Air
Power Range
100mW to 150W
Active Area Diameter
19 mm
Detector Coating
Broadband
Noise Equivalent Power (milliwatts)
3
Maximum Intermittent Power (watts)(<5 minutes)
150W @1 min., 100W @5 min., 50W @10 min.
Long-Pulse Joules Range (joules)
0.5 to 10
Wavelength Range (micrometers)
0.25 to 10.6 µm
Maximum Power Density (kilowatts per square centimeter)
6 kW/cm²
Maximum Energy Density (millijoules per square centimeter)
600 mJ/cm² 10 ns, 1064 nm
Response Time, Speed-up On, 0% to 95% (seconds)
3
Window
None
Diffuser
None
Calibration Uncertainty (%)(k=2)
±5
Power Linearity (%)
±1
Spectral Compensation Accuracy (%)
±1.5
Calibration Wavelength (nanometers)
10600
Cable Length (meters)
2.5